PCX-720 Troubleshooting This Troubleshooting Guide walks you step-by-step through the diagnosis of PCX-720 customer problems. It is important to follow these steps carefully, to ensure that the problem is diagnosed correctly, and therefore the appropriate repair action is taken. Getting Started 1.
Page 4
To Troubleshoot This Do the following: Issue If the footrest will not raise/lower at all, replace the footrest actuator. If that does not solve the problem, do one of the following: • For serial numbers CWC24130001 - My footrest will not raise/lower.
Page 5
Power Step 1: Check the power cord. 1. Check the entire length of the power cord for damage; if the power cord is damaged, replace the control module. Step 2: Check the wall outlet. 1. Plug a different device, such as a known working lamp, into the AC outlet into which the chair is plugged.
Page 6