• Do not use the ReliOn® Prime system at altitudes in excess of 10,000 ft
(3,048 m).
• Triglycerides up to 3,300 mg/dL do not significantly affect test results.
However, glucose values in specimens beyond this triglycerides level should
be interpreted with caution.
• Metabolites of icodextrin (maltose, maltotriose, and maltotetraose) do not
affect test results.
• Use only fresh capillary blood. Do not use serum or plasma or venous
whole blood.
IMPORTANT: Low or high blood glucose readings can indicate a potentially
serious medical condition. If your blood glucose reading is unusually low or
high, or you do not feel the way your readings indicate, repeat the test using
a new strip. If your result is still not consistent with your symptoms or if your
blood glucose result is less than 60 mg/dL or higher than 240 mg/dL, contact
your diabetes healthcare professional for advice.