Specifications - Fluke 62 MAX + User Manual

Hide thumbs Also See for 62 MAX +:
Table of Contents

Advertisement

Specifications

Temperature Range
Accuracy
(Calibration geometry with
ambient temperature 23 C
2 C)
Response Time (95 %)
Spectral Response
Emissivity
Temperature Coefficient
Shop for Fluke products online at:
62 MAX
-30 C to 500 C
(-22 F to 932 F)
0 C: 1.5 C or 1.5 % of
reading, whichever is
greater
( 32 F: 3 F or 1.5 % of
reading, whichever is
greater)
-10 C to <0 C: 2 C
( 14 F to <32 F: 4 F)
< -10 C: 3 C
(<14 F: 6 F)
<500 ms (95 % of reading)
8 to 14 microns
0.10 to 1.00
0.1 C/ C or 0.1 %/ C of reading (whichever is greater)
MyFlukeStore
www.
Infrared Thermometer
62 MAX +
-30 C to 650 C
(-22 F to 1202 F)
0 C: 1 C or 1 % of
reading, whichever is
greater ( 32 F: 2 F or
1 % of reading, whichever
is greater)
-10 C to <0 C: 2 C
( 14 F to <32 F: 4 F)
< -10 C: 3 C
(<14 F: 6 F)
<300 ms (95 % of reading)
1.877.766.5412
.ca
Specifications
7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents