Note
Under certain conditions, the OCP circuit may fail to clear because load demand occurs before the power
supply has time to build up the required output current capacity. In such cases, disable the output (press

Unregulated Operation

If the power supply goes into a mode of operation that is neither CV nor CC, the Unr annunciator will come on. An
unregulated condition limits the output current to a value that is safe for the supply. Some unregulated states occur so briefly
that they do not turn on the Unr annunciator. One condition that can cause a noticeable unregulated state is low ac line
voltage.

Saving and Recalling Operating States

Normal Operation

You can save programming time by storing up to 5 power supply operating states in nonvolatile memory. The programming
parameters that are saved are:
Output voltage Output current *OVP voltage
OCP state (on or off) Output state (enabled or disabled)
As an example, set up the following state:
Voltage = 4 V
Current = 1 A
OCP = on (OCP annunciator on) Output = Off (Dis annunciator on)
Save the above state to location 1 by pressing
Now set up the following state:
Voltage = 8 V
Current = 1.5 A
OCP = off (OCP annunciator off)
Save the above state to location 2 by pressing
Now restore the first state by pressing
. Note how the power supply is automatically programmed each time.

Turn-On Operation

Whenever you apply power to a new power supply it automatically turns on in a safe reset state with the following
parameters:
off
maximum
It is recommended that you leave the turn-on conditions as programmed. However, you may change them if you wish. To do
this, proceed as follows:
1.
Set up the power supply to the state you want when it is turned on.
60
Front Panel Operation
)before clearing the OCP circuit. After OCP is cleared, enable the power supply output.
OVP voltage = 5.5 V
OVP voltage = 8.5 V
Output = On (Dis annunciator off)
minimum
near zero
off
.
.
and verify the parameters. Restore the second state by pressing