8 Technical Specifications

Electrophoresis Power Supply EPS 601 User Manual 18-1130-19 Edition AD
Regulation
Output range
Programming range
Output resolution
Programming resolution
Accuracy
Line regulation
Load regulation
Ripple
Short term stability
Long term stability
Start current check
Sudden load change check
Ground leakage check
Output protection
Recovery after power failure
Maximum voltage, current and power with automatic cross-
over at preset limits
Voltage:
0–600 V DC
Current:
0–400 mA
Power:
0–100 W
Voltage:
6–600 V DC
Current:
1–400 mA
Power:
1–100 W
Time:
00:01–500 h
Volthour:
1–500 000 Vh
Voltage:
1 V
Current:
100 μA, 0–39.9 mA
1 mA, 40–400 mA
Power:
1W
Voltage:
1 V
Current:
1 mA
Power:
1 W
Time:
1 min, 00:01–99:59,
1h, 100–500 h
Volthour:
1 Vh, 1–9999 Vh,
100 Vh, 10.0–99.9 kVh
1 kVh, 100–500 kV
Voltage:
1%, ±3 V
Current:
1%, ±1 mA
Power:
2%, ±1 W
Timer:
0.1% ±1 min, 00:01–99:59 h
0.1%, ±10 min, 100–500 h
< 0.2%
< 1% at load change 10–90% of maximum load
< 1% at 600 V
< 0.2% /10 h after warm up
< 1% /year
Resistance not greater than 0.2 Mat 40 V
(current less than 200 μA). Can be disabled
Leakage not greater than 500 μA
Fully protected against any overload conditions
Duration < 8 s, The program continues automatically
Duration ≥ 8 s, The program continues after manual restart
Technical Specifications 8