Payne PA13NR018 Installation Instructions Manual page 6

Pa13nr split-system air conditioner for use with r-22 refrigerant
Table of Contents

Advertisement

NOTE:
These are not back-seating
valves.
It is not necessary
to
force the stem tightly
against the rolled lip.
The service valve
cap is a primary
seal for the valve
and must be
properly
tightened
to prevent
leaks.
Make
sure cap is clean
and
apply refrigerant
oil to threads
and sealing surface on inside of cap.
Tighten
cap finger tight and then tighten
additional
1/6 of a turn (1
wrench
flat) to properly
seat the sealing
surfaces.
I NOTE:
Using the service
ports (or gauge
set), release
the
I
nitrogen
pressure
from the system
before
attaching
vacuum
pump.
I.
EVACUATING
CONDENSER,
LINE
SET,
AND
INDOOR
COIL
PRODUCT
DAMAGE
HAZARD
Failure to follow this caution
may result in product
damage.
Never use the outdoor
unit compressor
as a vacuum
pump.
Doing
so may damage
the compressor.
Condenser,
line set, and indoor
coil should
be evacuated
using the
recommended
deep
vacuum
method
of 500
microns.
If deep
vacuum
equipment
is not available,
the alternate
triple evacuation
method
may be used by following
the specified
procedure.
If vacuum
must be interrupted
during
the evacuation
procedure,
always break vacuum
with dry nitrogen.
UNIT
OPERATION
HAZARD
Failure to follow this caution may result in improper unit
operation..
This unit is filled with a Nitrogen charge. Prior to starting
this unit,
evacuate
to 500 microns
and
charge with
refrigerant listed on rating label. Purge nitrogen before
connection to vacuum pump.
Deep Vacuum
Method
The deep
vacuum
method
requires
a vacuum
pump
capable
of
pulling
a vacuum
to 500 microns
and a vacuum
gauge capable
of
accurately
measuring
this vacuum
level. The deep vacuum
method
is the most
positive
way of assuring
a system
is free of air and
water.
Watch
the vacuum
gauge
as the
system
is pulling
down.
The
response
of the gauge is an indicator
of the condition
of the system
(refer to Fig. 7).
With no leaks in the system,
allow the vacuum
pump to run for 30
minutes
minimum
at the deep vacuum
level.
5000
4500
4000
3500
g3000
_2500
m
2000
1500-
VACUUM TIGHT
TOO WET
1000
500
TIGHT
DRY SYSTEM
0
1
2
3
4
5
6
7
MINUTES
Fig. 7 -
Deep Vacuum
Gauge
Response
and System
Conditions
A95424
Triple
Evacuation
Method
The triple evacuation
method
should
only
be used when
system
does not contain
any water
in liquid
form
and vacuum
pump
is
only
capable
of pulling
down
to 28 inches
of mercury
(711 mm
Hg). Refer to Fig. 8 and proceed
is as follows:
1. Pull system
down
to 28 inches
of mercury
(711 mm Hg)
and allow pump
to continue
operating
for an additional
15
minutes.
2. Close manifold
valves
or valve
at vacuum
pump
and shut
off vacuum
pump.
3. Connect
a nitrogen
cylinder
and regulator
to system
and fill
with nitrogen
until system
pressure
is 2 psig.
4. Close nitrogen
valve
and allow system
to stand for 1 hour.
During
this time, dry nitrogen
will diffuse
throughout
the
system
absorbing
moisture.
5. Repeat
this procedure
as indicated
in Fig. 8.
6. After the final evacuate
sequence,
confirm
there are no leaks
in the system.
If a leak is found,
repeat
the entire
process
after repair is made.
IEVAOUATE
I
IBREAKVACUUM
W ITH DRY NITROGEN
I
r- Tq
IEVACUATE I
IBREAK VACUUM WITHDRYNITROGEN
I
IEVACUATE I
ICHECK
FOR TIGHT'
DRY SYSTEM I
(IF IT HOLDS DEEP VACUUM)
I CHARGESYSTEM
I
Fig. 8 - Triple
Evacuation
Sequence
A95425
J.
GAUGE
PORTS
Check
for leaks
at the schrader
ports
and tighten
valve
cores
if
necessary.
Install plastic caps finger tight.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pa13nr024Pa13nr036Pa13nr030Pa13nr042Pa13nr048Pa13nr060

Table of Contents