Re-Installing The Bottom Panel - Hach SIGMA 900 Instrument Manual

Composite sampler
Hide thumbs Also See for SIGMA 900:
Table of Contents

Advertisement

Section 5

5.7 Re-installing the Bottom Panel

Figure 9
Tightening the Nuts on the Bottom Panel
Page 34
Re-installing the Bottom Panel
4. Place the controller, face down, on a cloth-covered workbench to prevent
scratching the front panel.
5. Inside the Center Section, remove the four screws that fasten the
Controller.
6. Lift the Center Section off the controller.
7. Remove the 17 screws from around the perimeter of the controller
back panel.
8. Carefully pull open the bottom panel and let the attached connectors
swing out of the way.
9. If necessary for the repair being performed, unplug the appropriate
connectors. Always note each connector location before unplugging
anything. Re-installing a plug into the wrong connector can cause
extensive damage.
Note: The front panel gasket has a light coating of grease to help assure a watertight
seal. Do not contaminate the grease or gasket area with dirt or debris during
servicing. Always replace the gasket if it is damaged or missing. Never
reassemble the case without the gasket properly installed.
Note: Always follow the procedure below when re-installing the 900 bottom panel.
Improper panel installation may result in damage to the instrument.
1. Hand tighten each nut until it makes contact with the panel.
2. Tighten the nuts in sequence shown in
3. Repeat the tightening procedure in the same sequence to 10 in.-lb
(1.13 N-m).
1
10
2
9
17
8
16
7
15
6
Figure 9
to 5 in.-lb (0.56 N-m).
11
3
12
4
13
5
14
8544Maintenance.fm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents