10
Triggers
4 Press the Slope softkey and select rising edge, falling edge, alternating
126
Digital channel (on mixed- signal oscilloscopes), D0 to the number of
digital channels minus one.
External — triggers on the rear panel EXT TRIG IN signal.
Line — triggers at the 50% level of the rising or falling edge of the AC
power source signal.
WaveGen — triggers at the 50% level of the rising edge of the
waveform generator output signal. (Not available when the DC or
Noise waveforms are selected.)
You can choose a channel that is turned off (not displayed) as the
source for the edge trigger.
The selected trigger source is indicated in the upper- right corner of the
display next to the slope symbol:
1 through 4 = analog channels.
D0 through Dn = digital channels.
E = External trigger input.
L = Line trigger.
W = Waveform generator.
edges, or either edge (depending on the selected source). The selected
slope is displayed in the upper- right corner of the display.
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide