5.6.1 HF Gain Input 1.

Proceed as follows to do the HF Gain Input 1&2 calibration:
1. Press
2. Connect the test tool to the 5500A as shown in Figure 5-3. Do NOT use a 50Ω
3. Set the 5500A to source a 1 kHz fast rising edge square wave (Output SCOPE,
4. Set the 5500A in operate (OPR).
5. Press
6. Wait until the display shows calibration status
7. Press
8. Set the 5500A to source a 1 kHz square wave (Output SCOPE, MODE wavegen,
9. Press
10. Press
11. Wait until the display shows calibration status
12. Press
13. When you are finished, set the 5500A to Standby.
14. Continue at Section 5.6.2.
5-8
to select the first calibration step in Table 5-1 (
termination!
Figure 5-3. HF Gain Calibration Input Connections
MODE edge) to the first calibration point in Table 5-1.
to start the calibration.
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points in Table 5-1.
WAVE square), to the first calibration point in Table 5-2.
to select the first step in Table 5-2.
to start the calibration.
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points Table 5-2.
HFG & FI AB (CL 0600):
.
READY
.
READY
)
ST8097.WMF