Description Of The Item "Data Output Mode - A&D FZ-120iWP Instruction Manual

Dust and waterproof precision balances fz-iwp series/fx-iwp series
Hide thumbs Also See for FZ-120iWP:
Table of Contents

Advertisement

9-5 Description of the Item "Data Output Mode"

The parameter setting of the "Data output mode ( prt )" applies to the performance when the data is
transmitted using the RS-232C serial interface.
Key mode
When the
PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data and the display blinks one time.
Required setting
Auto print modes A and B
When the displayed value is stable and the conditions of "Auto print polarity", "Auto print difference" and
reference value are met, the balance outputs the weight data.
When the
PRINT key is pressed with the stabilization indictor turned on, the balance outputs the data
and the display blinks one time.
Mode A: Required setting
Example
Mode B: Required setting
Example
Stream mode
The balance outputs the weight data continuously regardless of the display condition. The display does
not blink in this mode.
Required setting
Example
Caution
The balance may not transmit the data completely at the specified refresh rate and baud rate. Set
the baud rate higher.
Key mode B
When the
PRINT key is pressed, the balance outputs the weight data, regardless of the display
condition. The display does not blink in this mode.
Key mode C
When the
PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on and
outputs the data. The balance blinks one time in this mode.
dout
prt 0
dout
prt 1
dout
ap-p
dout
ap-b
"For outputting the weight value each time a sample is added, with
"ar-d" set to "1" (to adjust zero after the data is output)."
dout
prt 2
dout
ap-p
dout
ap-b
"For outputting the weight value while a sample is added."
dout
prt 3
ba5fnc
5pd
5if
bp5
"For monitoring data on a computer"
42
Key mode
Auto print mode A (reference = zero)
Auto print polarity
Auto print difference
Auto print mode B (reference = last stable value)
Auto print polarity
Auto print difference
Stream mode
Display refresh rate
Baud rate

Hide quick links:

Advertisement

Table of Contents
loading