GE MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT Operation And Maintenance Manual

Ge osmonics milleniumtm reverse osmosis unit operation and maintenance manual
Table of Contents

Advertisement

GE OSMONICS
MILLENIUM™ REVERSE OSMOSIS UNIT
Operation
and Maintenance Manual
SUPERCEDES P/N 14693 REV. D

Advertisement

Table of Contents
loading

Summary of Contents for GE MILLENIUM GE OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT

  • Page 1 GE OSMONICS MILLENIUM™ REVERSE OSMOSIS UNIT Operation and Maintenance Manual SUPERCEDES P/N 14693 REV. D...
  • Page 3: Table Of Contents

    MILLENIUM REVERSE OSMOSIS UNIT Installation, Operation and Maintenance Manual CHAPTER ONE: GENERAL INFORMATION Section The Manual Safety Summary Applications Contraindications Environmental Considerations Theory of Operation Fluid Components Monitors and Controls Optional Equipment 1.10 Specifications 1.10.1 Inlet Water Requirements 1.10.2 Flow Rates 1.10.3 Electrical Requirements 1.10.4...
  • Page 4 CHAPTER FOUR: ROUTINE MAINTENANCE Section Routine Maintenance of Millenium™ RO System 4.1.1 Daily / Per Treatment Maintenance 4.1.2 Weekly Maintenance 4.1.3 Monthly Maintenance 4.1.4 Quarterly Maintenance 4.1.5 Semi-Annual Maintenance 4.1.6 Annual Maintenance Specific Maintenance Procedures and Instructions 4.2.1 Motor Lubricant Procedure 4.2.2 Leak Repair 4.2.3...
  • Page 5 CHAPTER FIVE: TROUBLESHOOTING Table of Content Section Unit Fails to Run Unit Runs Intermittently Motor Runs Hot Water Produced While Unit is Off Low Pump Pressure Pressure Gauge Vibrates Erratically Excess Product Pressure (too high) Product Flow Too Low Product Flow Too High 5.10 Leaks 5.11...
  • Page 6 This page intentionally left blank 14693e 31 March 2003...
  • Page 7: Chapter One

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER ONE: GENERAL INFORMATION 14693e – 31 Mar 03 1 - 1 General Information...
  • Page 8 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e – 31 Mar 03 1 - 2 General Information...
  • Page 9: The Manual

    The Manual This manual has been prepared to provide the operator with information and instructions regarding the installation, use, maintenance and troubleshooting of the GE Osmonics Millenium™ Reverse Osmosis System. CAUTION: When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.
  • Page 10: Applications

    GE Osmonics systems are not designed, sold, or intended for use in producing water for injection Furthermore, GE Osmonics systems are not intended to be used outside of the device specifications and limitations, as outlined in this manual and other related materials.
  • Page 11: Environmental Considerations

    NOISE: The RO unit does not generate significant noise levels. Hard walls may reflect noise and will make the RO seem louder. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics ° ° F (25 C) feed temperature.
  • Page 12: Theory Of Operation

    The following paragraphs detail the function of these fluid control components, which are used within the RO. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 6 General Information...
  • Page 13: Fluid Components

    "flush" from the full open (flush) to just before the fully closed (operate) position. CAUTION: Excessive “chattering” may be eliminated by slowly moving the valve from the full flush position towards “OPERATE”. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 7 General Information...
  • Page 14 When drawing disinfectants or cleaning solutions, do not allow the RO to draw air. This may cause irreparable damage to the RO pump. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 8 General Information...
  • Page 15: Monitors And Controls

    RO. The port has a check valve that closes when there is no hose connected to the fitting. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics The inlet water switch must be ON for the 1 - 9...
  • Page 16 The Alarm Mute switch will mute the audible alarm for 30 - 45 seconds. A red indicator lamp is displayed on the switch when the mute switch has been pressed. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 10 General Information...
  • Page 17: Optional Equipment

    1.9 Optional Equipment GE Osmonics offers a number of optional items to adapt the equipment to meet specific needs. Options are available to improve portability, and to improve the quality of the feedwater to the RO. Frequently requested options are briefly described in the following paragraphs.
  • Page 18: Specifications

    Waste Flow: .84 GPM (3.2 LPM) ( 1.10.3 Electrical Requirements Motor Voltage Hertz Amps Phase 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics Minimum 2.0 GPM 1.7°C (35°F) 30 PSI 190 PSI 1 PSI ° F (25 °...
  • Page 19: Dimensions

    1.11 Disposal Disposal of this product or parts must be carried out according with local disposal codes. 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics Vertical Cabinet 25 ½ inches 15 inches 8 inches 95% minimum at initial testing at factory.
  • Page 20: Electromagnetic Interference

    Contact your Facility Equipment Technician. If the technician is unable to help then… Call GE Osmonics Technical Service Department at (800) 633-3080 or (206) 395-2200. Technicians are available for all calls between 7:00 a.m. and 4:30 p.m. PST, Monday through Friday.
  • Page 21: Return Goods Authorization (Rga)

    Return Goods Authorization (RGA) Procedure If you wish to return goods for, warranty evaluation and/or credit please have your original sales order, invoice and device serial number available when you call GE Osmonics. Call GE Osmonics at (800)-633-3080 and request Technical Service. A GE...
  • Page 22: Symbols And Abbreviations

    Granular Activated Carbon Hertz Horse Power lbs. Pounds Light Emitting Diode Liters Per Day Liters Per Minute Meter mg/L. Milligrams Per Liter Milliliters 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 16 General Information...
  • Page 23 9 on the Main Terminal Barrier Strip) Thin Film Total Dissolved Solids Underwriters Laboratory Volts Alternating Current Water for Injection Weight Volts Alternating Current 14693e – 31 Mar 03 M illenium™ RO System by GE Osmonics 1 - 17 General Information...
  • Page 24 M illenium™ RO System by GE Osmonics NOTES: 14693e – 31 Mar 03 1 - 18 General Information...
  • Page 25: Installation Guidelines And Initial System Start Up

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER 2: INSTALLATION GUIDELINES AND INITIAL SYSTEM START-UP 14693e - 31 Mar 03 2 - 1 Installation Guidelines and Instructions...
  • Page 26 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 2 - 2 Installation Guidelines and Instructions...
  • Page 27: Installation Considerations

    25°) will produce more permeate, and will consume more water. The water may also require some pretreatment to ensure maximum performance and suitability. GE Osmonics recommends that feed water be chlorine/chloramine free and solids above 5 microns in size be filtered out.
  • Page 28 200 PSI. If 200 PSI is exceeded, adjust pump pressure on pump head located inside the cabinet. See instructions in Figure 1 below: 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 2 - 4 Installation Guidelines and Instructions...
  • Page 29 • Microbiological Culture (colony count) results • Endotoxin (pyrogen) level testing (if applicable) The Millenium™ RO is now ready for use. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 2 - 5 Installation Guidelines and Instructions...
  • Page 30: Product Water Analysis Procedure

    AAMI Analysis. Chlorine and chloramine dissipate from the water sample over time and reported results will be inaccurate. NOTES: 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 2 - 6 Installation Guidelines and Instructions...
  • Page 31 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 2 - 7 Installation Guidelines and Instructions...
  • Page 32: System Operation

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER 3: SYSTEM OPERATION 14693e - 31 Mar 03 System Operation...
  • Page 33 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 System Operation...
  • Page 34: Daily Start-Up Procedure

    The Millenium™ RO is now ready for connection to the dialysis machine. Shut Down Procedure Turn off the RO pump power switch. Turn off the inlet water valve switch. Turn off the inlet water supply. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics System Operation...
  • Page 35 RO should be stored in an approved disinfectant or storage solution (Not ‘RENALIN®’ or other type oxidizing products). Refer to the disinfection or storage procedure in this manual. Disconnect from the utilities and store the RO. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics System Operation...
  • Page 36 NOTES: 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics System Operation...
  • Page 37 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 System Operation...
  • Page 38: Chapter Four: Routine Maintenance

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER FOUR: ROUTINE MAINTENANCE 14693e - 31 Mar 03 4- 1 Routine Maintenance...
  • Page 39 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 4- 2 Routine Maintenance...
  • Page 40: Section 4.1 Routine Maintenance Of Millenium™ Ro System

    Routine Maintenance of Millenium™ RO System GE Osmonics Millenium™ RO systems have been designed to operate with a minimum of operator attention. Like all mechanical systems they will operate longer and with less trouble when operator maintenance is performed regularly. Operator maintenance on a...
  • Page 41: Weekly Maintenance

    Check the quick-disconnect fittings and hoses. They should seat securely and fluid should shut off when they are disconnected. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics Refer to the disinfection procedure in this 4- 4...
  • Page 42: Semi-Annual Maintenance

    Check the switches and lights for correct operation. Check all pre RO devices for correct operation. Lubricate the motor. Refer to the lubricating instructions in this chapter. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 5 Routine Maintenance...
  • Page 43: Specific Maintenance Procedures And Instructions

    Check again for leaks. (Over tightening can crack the female fitting). If the leak continues, it may be necessary to replace the male threaded fitting and/or the female threaded fitting. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 6 Routine Maintenance...
  • Page 44 M illenium™ RO System by GE Osmonics Tubing connections: 14693e - 31 Mar 03 4- 7 Routine Maintenance...
  • Page 45 Re-pressurize the system (if applicable) and inspect for leaks. If the leak is not resolved, start over or simply replace the entire tube fitting or tube. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 8 Routine Maintenance...
  • Page 46: Product Recirculation Check Valve Test

    Reapply pressure in opposite direction of flow and recheck pressure and for seepage. If pressure does not drop and no seepage is found, valve is functioning correctly. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 9 Routine Maintenance...
  • Page 47: Millenium™ Cleaning Indications

    Regular maintenance cleaning is the best assurance of long membrane service. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics A good preventative 4- 10 Routine Maintenance...
  • Page 48: Pre Cleaning Preparation

    Materials Required: Liqui-Clean 403™ – (Low pH liquid) - GE Osmonics Part No. 1228653 (1 Gal) Liqui-Clean 211™ – (High pH liquid) - GE Osmonics Part No. 1228791 (1 Gal) • high and low pH cleaner •...
  • Page 49: Cleaner Recirculation And Dwell Time

    Using the disinfect switch, turn the RO on and flush approximately 1/2 gallon of the solution to the drain. This prevents fouling the cleaning solution with the first 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 12 Routine Maintenance...
  • Page 50: Post Cleaning Rinse Procedure

    Repeat the cleaning procedure with the other pH cleaner. recommends using a low pH cleaner first, followed by a high pH cleaner after rinsing thoroughly. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 13 Routine Maintenance GE Osmonics...
  • Page 51 Reconnect the product line to its original connection. Remove the “Do Not Use/Contains Cleaner” warning label. The Millenium™ RO system is now ready for use. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 14 Routine Maintenance...
  • Page 52: Biosan™ Pre Cleaning Preparation

    BIOSAN™" to prevent use of RO until properly rinsed. CAUTION: Water to be used must be chlorine free or damage to the RO membranes may result. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 15 Routine Maintenance Neither...
  • Page 53: Biosan™ Recirculation And Dwell Time

    Allow the BIOSAN™ to dwell in the RO for a minimum of one hour. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 16 Routine Maintenance...
  • Page 54: Biosan™ Post Cleaning Rinse Procedure

    Reconnect the product line to its original connections. Remove the 'Do Not Use/Contains BIOSAN™' warning label. The Millenium™ RO System is now ready for use. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 17 Routine Maintenance...
  • Page 55: Millenium™ Disinfection Indications

    • “DO NOT USE” warning labels or signs (not provided by Osmonics) • cleaning / disinfect log sheets • most current log sheets for water quality comparison • proper protective body wear WARNING: 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 18 Routine Maintenance...
  • Page 56: Formaldehyde Pre Disinfection Procedure

    Do not allow air to be drawn into the RO system, as this will damage the pump. Test for the presence of formaldehyde in the product line (the RO must be running). Release the disinfect switch. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 19 Routine Maintenance...
  • Page 57: Formaldehyde Post Disinfection Rinse Procedure

    Check the product line for residual disinfectant with an appropriate test kit. If the test is positive, continue to rinse until a negative test is obtained. Disconnect the product rinse line and store it. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 20 Routine Maintenance...
  • Page 58 Success of the disinfection procedure may be ascertained by taking a water culture after completion. See the Product Water Culture Procedure in this chapter. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 21 Routine Maintenance...
  • Page 59: Renalin®" Disinfection Procedure

    • “DO NOT USE” warning labels or signs (not provided by Osmonics) • cleaning / disinfect log sheets • most current log sheets for water quality comparison 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 22 Routine Maintenance...
  • Page 60: Renalin® Pre-Disinfection Preparation

    CAUTION: When ‘RENALIN®’ is to be used as a disinfectant, several preparatory steps are required by GE Osmonics as well as the membrane manufacturer. Failure to comply with these instructions, which incorporate instructions from the membrane manufacturer, will jeopardize the system performance and warranty.
  • Page 61: Renalin® Disinfection And Dwell Time

    Turn on the inlet water valve switch and allow the system to rinse for 2 minutes. Turn the RO pump power switch on and run the RO for 10 minutes. CAUTION: 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 24 Routine Maintenance...
  • Page 62 Disconnect the product rinse line and store it. Remove the 'Do Not Use/Contains Disinfectant’ warning label. The Millenium™ RO System is now ready for use. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 25 Routine Maintenance...
  • Page 63: Product Water Culture

    C) immediately and assayed within a 24 hour period. Obtain the samples wearing long sleeves and a mask to prevent contamination of samples, use a “mid-stream” “clean catch” type procedure. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 26 Routine Maintenance...
  • Page 64: Product Water Culture Procedure

    • Any other pertinent information or procedures your facility or lab requires The samples should be assayed within 30 minutes of obtaining or refrigerated at 5° C immediately and assayed within a 24-hour period. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 27 Routine Maintenance...
  • Page 65: System Storage

    After storage, the RO unit must be completely rinsed of formaldehyde, and a bacterial culture performed and interpreted before use. Perform AAMI analysis if it has been one year or more since last done. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 28 Routine Maintenance...
  • Page 66: Biosan™ Storage Procedure

    After storage, the RO unit must be completely rinsed of BIOSAN™, and a bacterial culture performed and interpreted before use. Perform AAMI analysis if it has been one year or more since last done. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 4- 29 Routine Maintenance...
  • Page 67 M illenium™ RO System by GE Osmonics NOTES: 14693e - 31 Mar 03 4- 30 Routine Maintenance...
  • Page 68: Chapter Five: Troubleshooting

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER FIVE: TROUBLESHOOTING 14693e - 31 Mar 03 5- 1 Troubleshooting...
  • Page 69 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 5- 2 Troubleshooting...
  • Page 70: Table Of Content

    TDS Value High 5.13 RO Fails to Draw Disinfectant 5.14 RO Pump Leaks 5.15 TDS Value fluctuates 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics Table of Contents TITLE 5- 3 PAGE 5-10 5-11 5-11 5-11...
  • Page 71 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 5- 4 Troubleshooting...
  • Page 72: Ro Fails To Run

    See 5.5. RO Runs Intermittently Loose electrical connection. a. Verify that all wiring connections are secure. b. Verify that plug is securely connected to outlet. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 5 Troubleshooting...
  • Page 73: Motor Runs Hot

    Remove valve and clear obstruction. c. Replace valve. Fault inlet water valve switch. a. Verify switch is in OFF position. b. Check switch with ohm meter, replace switch. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 6 Troubleshooting...
  • Page 74: Low Pump Pressure

    Inlet regulator faulty. a. Replace regulator. Excess Product Pressure (too high) Inlet pressure regulator improperly adjusted. a. Decrease pressure regulator to adjust maximum product setting. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 7 Troubleshooting...
  • Page 75: Product Flow Too Low

    Regenerate softener. d. Check supply water for hardness and conductivity. Product recirculation check valve failed. a. Refer to Maintenance Procedures for Check Valve Test. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 8 Troubleshooting...
  • Page 76: Product Flow Too High

    General information: Any leak from any part of the system should be corrected as soon as possible. necessitating expensive repairs. Refer to the Maintenance Procedures for Leak Repair. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics Left unattended, leaks can cause secondary damage 5- 9 Troubleshooting...
  • Page 77: Low Percent Rejection

    Change in feed water TDS. a. Verify feed and product water quality with independent meter. b. Softener problem, rinsing out high levels of salt. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 10 An apparent Troubleshooting...
  • Page 78: Tds Value High

    Softener problem, rinsing out high levels of salt. Product recirculation check valve failed. a. Refer to Maintenance Procedures for Check Valve Test. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics If the product water quality is poor, investigate the 5- 11 An apparent...
  • Page 79: Ro Fails To Draw Disinfectant

    Mechanical seal failure (leaks out vent hole). a. Replace pump. 5.15 TDS Value Fluctuates Product recirculation check valve failed. a. Refer to Maintenance Procedures for Check Valve Test. 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 5- 12 Troubleshooting...
  • Page 80 M illenium™ RO System by GE Osmonics NOTES: 14693e - 31 Mar 03 5- 13 Troubleshooting...
  • Page 81 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 5- 14 Troubleshooting...
  • Page 82: Chapter Six: Drawings

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER SIX: DRAWINGS 14693e - 31 Mar 03 6 - 1 Drawings...
  • Page 83 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 6 - 2 Drawings...
  • Page 84 M illenium™ RO System by GE Osmonics Millenium RO (Component Layout) 14693e - 31 Mar 03 6 - 3 Drawings...
  • Page 85 M illenium™ RO System by GE Osmonics Millenium RO (Disinfect View) 14693e - 31 Mar 03 6 - 4 Drawings...
  • Page 86 M illenium™ RO System by GE Osmonics Millenium RO (Flow Schematic) 14693e - 31 Mar 03 6 - 5 Drawings...
  • Page 87 M illenium™ RO System by GE Osmonics Millenium RO (Electrical Schematic, High Voltage, 50/60Hz) 14693e - 31 Mar 03 6 - 6 Drawings...
  • Page 88 M illenium™ RO System by GE Osmonics Millenium RO (Electrical Schematic, Low Voltage, 50/60Hz) 14693e - 31 Mar 03 6 - 7 Drawings...
  • Page 89 ELECTRICAL CONTROL PANEL 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 21080 20204 20205 20203 30005 ELECTRICAL CONTROL BOX Millenium RO (Electrical Layout) 6 - 8 12312 12310 30010 1236062 (60Hz) 1234687 (50Hz) 12315 (60HZ) 12315−1 (50HZ)
  • Page 90 M illenium™ RO System by GE Osmonics 40569 1234678 Millenium RO (Input Manifold 50/60Hz) 14693e - 31 Mar 03 6 - 9 Drawings...
  • Page 91 M illenium™ RO System by GE Osmonics 14765 40218 12131 40569 40564 40602 20178 Millenium RO (Main Manifold) 14693e - 31 Mar 03 6 - 10 Drawings...
  • Page 92 M illenium™ RO System by GE Osmonics 14738 40976 40976 40976 40218 40976 40976 40570 40218 Millenium RO (Waste Manifold) 14693e - 31 Mar 03 6 - 11 Drawings...
  • Page 93 M illenium™ RO System by GE Osmonics OS1151581 DETAIL ’A’ (NOT TO SCALE) Millenium RO (ROMA Assembly) 14693e - 31 Mar 03 6 - 12 Drawings...
  • Page 94 M illenium™ RO System by GE Osmonics 50112 14493 41566 41566 42014 TO HOLE IN RO CABINET Millenium RO (Flush Valve Assembly) 14693e - 31 Mar 03 6 - 13 Drawings...
  • Page 95 5. REPLACE WASHER (IF IT CAME OFF) AND ACORN NUT. 6. TURN ON WATER AND POWER AND THEN CHECK PRESSURE READING. REPEAT AS NECESSARY. PUMP (Pump Pressure Adjustment Procedure) 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics PRESSURE ADJ. SCREW WASHER ACORN NUT Millenium RO...
  • Page 96 M illenium™ RO System by GE Osmonics “Push-In” Fitting Instructions 14693e - 31 Mar 03 6 - 15 Drawings...
  • Page 97 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 6 - 16 Drawings...
  • Page 98 M illenium™ RO System by GE Osmonics NOTES: 14693e - 31 Mar 03 6 - 17 Drawings...
  • Page 99 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 6 - 18 Drawings...
  • Page 100: Chapter Seven: Spare Parts List

    M illenium™ RO System by GE Osmonics Millenium RO System CHAPTER SEVEN: SPARE PARTS LIST 14693e - 31 Mar 03 7 - 1 Spare Parts List...
  • Page 101 M illenium™ RO System by GE Osmonics This page intentionally left blank. 14693e - 31 Mar 03 7 - 2 Spare Parts List...
  • Page 102 Pump, 100 GPH, SS, W/Regulator 50041 Pump-to-Motor Coupling 1234489 Regulator, Product Pressure Adjustment 12131 Sensor Probe, Total Dissolved Solids, Feed, Product 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics Millenium RO System 7 - 3 Spare Parts List...
  • Page 103 Hardness Test Kit 14604 Silt Density Index (SDI) Test Kit 11656 Total Chlorine Test Kit 70001 Water Test Kit, AAMI Standards OPTIONAL 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 7 - 4 Spare Parts List...
  • Page 104 Filter, 5 Micron, 9 7/8" 40698 Filter, Carbon, 9 7/8" 14924 Kit, Mil, Prefilter Install (Mounts prefilter to RO or to wall. Includes all necessary parts.) 14693e - 31 Mar 03 M illenium™ RO System by GE Osmonics 7 - 5 Spare Parts List...
  • Page 105 M illenium™ RO System by GE Osmonics NOTES 14693e - 31 Mar 03 7 - 6 Spare Parts List...
  • Page 107 Call (253) 395-2200 for additional information, (800) 633-3080 in the U.S., or visit www.osmonics.com Corporate Headquarters 5951 Clearwater Drive Minnetonka, MN 55343-8995 (952) 933 2277 Phone (952) 933-0141 Fax Kent Operations 7848 South 202 Kent, WA 98032-1345 (253) 395-2200 Phone (253) 395-2363 Fax...