Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INSTrument:COUPle[:TRIGger]?
This query returns the currently coupled output. Returns "ALL", "NONE", or a list.
If any output is not coupled, "NONE" is returned. If all of three outputs are coupled,
"ALL" is returned. If a list of outputs is coupled, the list is returned.
Measurement Commands
MEASure:CURRent[:DC]? [{P6V | P25V | N25V}]
This command queries the current measured at the output terminals of the power
supply. The physical outputs of measurement are specified by the output identifier. If
any output identifier is not specified, the current of the currently selected output is
returned.
MEASure[:VOLTage][:DC]? [{P6V | P25V | N25V}]
This command queries the voltage measured at the output terminals of the power
supply. If any output identifier is not specified, the voltage of the currently selected
output is returned.
78