Using Triggering to Start or Stop Scope Record
To record an electrical event that causes a fault, it might
be useful to start or stop recording on a trigger signal:
Start on trigger to start recording; recording stops when
the deep memory is full
Stop on trigger to stop recording.
Stop when untriggered to continue recording as long as
a next trigger comes within 1 division in view all mode.
For the models 190-xx4 the signal on the BNC input that
has been selected as trigger source must cause the
trigger.
For the models 190-xx2 the signal applied to the banana
jack inputs (
EXT TRIGGER (in)
trigger. The trigger source is automatically set to
(external).
To set up the test tool, continue from step 3 of the
previous section:
4
Apply the signal to be recorded to the BNC
input(s).
5
Stop recording to unlock the
F1
OPTIONS... softkey
Shop for Fluke products online at:
Recording Scope Waveforms In Deep Memory (Scope Record)
). signal must cause the
Ext.
MyFlukeStore
www.
Using The Recorder Functions
6
Open the
F2
menu.
7
Jump to the Mode: field, select
on Trigger... (models 190-xx4) or
ENTER
on Ext. (models 190-xx2) to open
the
START SINGLE SWEEP ON
TRIGGERING
SWEEP ON EXT
8
Select one of the Conditions:
and accept the selection.
ENTER
For external triggering (190-xx2) continue at step 9.
1.877.766.5412
.com
2
RECORDER OPTIONS
or the
START SINGLE
. menu.
47