18
Save/Recall (Setups, Screens, Data)
4 Finally, press the Press to Save softkey.
When saving screen images, the oscilloscope uses the last menu visited before pressing
N O T E
the [Save/Recall] key. This lets you save any relevant information within the softkey menu
area.
To save a screen image showing the Save/Recall Menu at the bottom, press the
[Save/Recall] key twice before saving the image.
You can also save the oscilloscope's display image using a web browser. See
N O T E
Image"
See Also
•
To save CSV, ASCII XY, or BIN data files
Data files can be saved to an external USB storage device.
1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select
278
•
Setup Info — setup information (vertical, horizontal, trigger, acquisition,
math, and display settings) is also saved in a separate file with a
TXT extension.
•
Invert Grat — the graticule in the image file has a white background
instead of the black backgound that appears on- screen.
Graticule Not Inverted
Palette — lets you choose between Color or Grayscale images.
•
A message indicating whether the save was successful is displayed.
on page 325 for details.
"Adding an Annotation"
CSV data (*.csv), ASCII XY data (*.csv), or Binary data (*.bin).
Graticule Inverted
on page 312
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
"Get