C. Replacing the Fuse (see figure 9)
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
To replace the Meter's fuse:
1.
Turn the Meter power off and remove all connections
from the terminals.
2.
Remove the screw from the battery compartment, and
separate the battery compartment from the case bottom.
3.
Remove the screws from the case bottom, and separate
the case top from the case bottom.
4.
Remove the fuse by gently prying one end loose, then
take out the fuse from its bracket.
5.
Install ONLY replacement fuses with the identical type
and specification as follows and make sure the fuse is
fixed firmly in the bracket.
Fuse 1: 0.315A, 250V, fast type fuse, 5x20 mm.
6.
Rejoin the battery compartment and the case top, and
reinstall the screw.
7.
Rejoin the case bottom and case top, and reinstall the
screws.
Replacement of the fuses is seldom required. Burning of
a fuse always results from improper operation.
This operating manual is subject to change without notice.