Agilent Technologies 6012B Service Manual page 109

Autoranging dc power supply
Hide thumbs Also See for 6012B:
Table of Contents

Advertisement

To check
CV
Mode
proceed as follows:
Using test set-up, Figure A-13, connect to end of 2K Ω resistor to J3-36.
a.
b. Turn on power supply.
c.
Using "Display Setting'' set voltage and current or power supply for 1 volt and 1 amp.
d. DVM should read between 0 to 0.4 volts.
e.
Turn off power supply and short to output terminals.
f.
Turn on power supply.
g.
DVM should read approximately 5Vdc.
To check
CC
Mode
proceed as follows:
Using test set-up, Figure A-13, connect top end of 2K Ω resistor to J3-35.
a.
b. Turn on power supply.
c.
Using "Display Settings'' set voltage for 1 volt and current for 1 Amp.
d. DVM should read = 5Vdc.
e.
Turn off power supply and short the output terminals.
f.
Turn on power supply.
g.
DVM should read between 0 to 0.4Vdc.
To check
OVERVOLTAG proceed as follows:
Using test set-up, Figure A-13, connect top end of 2K Ω resistor to J3-17.
a.
b. Turn "OVP Adjust" fully clockwise and voltage control fully counter clockwise.
c.
Open power supply output terminals and turn on power.
d. DVM should read approximately 5Vdc.
e.
Press 'Display Settings" and increase voltage control for 15Vdc output.
f.
Turn OVP Adjust'' counterclockwise until supply goes into overvoltage.
g.
DVM should read between 0 and 0.4Vdc.
h.
Turn "OVP Adjust" fully clockwise and turn off input power for 5 seconds.
i.
Turn on input power and DVM should read approximately 5Vdc.
To check
OUTPUT
UNREGULAT
Using test set-up, Figure A-13, connect to end of 2K Ω to J3-18.
a.
Figure A-13. Troubleshooting Status Indicators
E
ED
proceed as follows:
109