Agilent Technologies 6012B Service Manual page 10

Autoranging dc power supply
Hide thumbs Also See for 6012B:
Table of Contents

Advertisement

TYPE
Oscilloscope
Isolation Transformer
RMS Voltmeter
Logic Pulser
Multimeter
CC PARD Test
Current Probe
Electronic Load*
CC PARD Test
Resistive Load
10
Table 2-1. Test Equipment Required
REQUIRED CHARACTERISTICS
Sensitivity: 1mV
Bandwidth: 20MHz & 100MHz
Input: differential, 50Ω & 10MΩ
100VA 4KVA minimum
True rms, 10MHz bandwidth
Sensitivity: 1 mV
Accuracy: 5%
4.5 to 5.5Vdc @ 35mA
Resolution: 100nV
Accuracy: 0.0035%, 6½ digit
No saturation at:
6010A 20Adc
6011A 100Adc
6012B 51Adc
6015A 51Adc
Bandwidth: 20Hz to 20MHz
Power range: 1000 watts
Open and short switches
6010A
Voltage range: 200Vdc
Current range: 20Adc
6011A
Voltage range: 30Vdc
Current range: 120Adc
6012B
Voltage range: 65Vdc
Current range: 55Adc
6015A
Voltage range: 200Vdc
Current range: 5Adc
Value:
6010A
3.5 ohms >1000W
Accuracy: 1%
6011A
0.058 ohms >1000W
Accuracy 1%
6012B
0.4 ohms >1000W
Accuracy: 1%
Rheostat or Resistor Bank
USE
RECOMMENDED MODEL
P,T
Agilent 1740A
T
P
Agilent 3400A
T
Agilent 546A
P,A,T
Agilent 3456A
P
Tektronix P6303
Probe/AM503 Amp/
TM500 Power Module
P,A
Transistor Devices Model
DLP 130-50-2500
DLR-400-15-2500
DLP 50-150-3000
DLP 130-50-2500
DLR-400-15-2500
P,A