Fluke 705 Instruction Sheet page 4

Loop calibrator
Hide thumbs Also See for 705:
Table of Contents

Advertisement

Sourcing mA
Use source mode whenever you need to supply current into a
passive circuit. Insert the test leads into the terminals as shown
below.
A path must exist for current to flow between the output
and mA terminals, otherwise the display will indicate
an overload (OL) when you set an output value.
Source
Press to step output
up/down in 25 %
steps.
Press to toggle
Source or Simulate.
Note
705
LOOP CALIBRATOR
MEASURE
OFF
SOURCE
VDC
SIMULATE
Loop Power
mA
SpanCheck
+
COM
or
Press to scroll
up/down 0.01 mA.
Hold down to scroll
faster.
For 0.001 mA steps,
see "Pushbutton
Functions."
Press simultaneously
once for slow ramp
twice for fast ramp
three times for
stepped ramp.
rt04f.eps