Fluke 771 Calibration Manual

Milliamp process clamp meter
Hide thumbs Also See for 771:

Advertisement

Quick Links

®
771
Milliamp Process Clamp Meter
Calibration Manual
November 2007
© 2007 Fluke Corporation, All rights reserved. Printed in USA. Product specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 771

  • Page 1 ® Milliamp Process Clamp Meter Calibration Manual November 2007 © 2007 Fluke Corporation, All rights reserved. Printed in USA. Product specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for 3 years (one year for cable and clamp) from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction......................1 Contacting Fluke....................2 Safety Information ..................... 2 Symbols ......................3 Specifications..................... 4 Getting Acquainted with the Meter..............5 Maintenance....................... 6 Cleaning the Meter ..................6 Battery Replacement ..................6 Performance Tests....................8 Required Equipment..................8 Testing the Batteries ..................
  • Page 4 Calibration Manual...
  • Page 5 List of Tables Table Title Page Symbols........................3 Required Equipment....................8 Accuracy Tests ....................... 10 Error Messages....................... 12 Replaceable Parts ....................15...
  • Page 6 Calibration Manual...
  • Page 7 List of Figures Figure Title Page The 771 Milliamp Process Clamp Meter ............... 5 Changing the Batteries ................... 7 Display Test ......................9 Accessing the Calibration Button................12 Phase Adjustment....................14 Replaceable Parts ....................16...
  • Page 8 Calibration Manual...
  • Page 9: Introduction

    The Meter allows closed-case calibration using reference sources. It measures the reference signals, calculates the correction factors, and stores them in memory. The instrument should be calibrated after repair, or if it fails a performance test. The 771 Calibration Manual provides the following information: Precautions and safety information Specifications...
  • Page 10: Contacting Fluke

    1-800-363-5853 in Canada +31-402-675-200 in Europe +81-3-3434-0181 in Japan +65-738-5655 in Singapore +1-425-446-5500 from anywhere in the world Or, visit Fluke’s Web site at www.fluke.com To register your product, visit http://register.fluke.com Safety Information In this manual, a Warning identifies conditions and actions that pose hazard(s) to the user.
  • Page 11: Symbols

    Battery Low Battery Conforms to relevant European Union directives DC (Direct Current) Do not dispose of this product as unsorted municipal waste. Go to Fluke’s web site for recycling information. Conforms to relevant Australian standards N10140 Conforms to relevant Canadian and US standards...
  • Page 12: Specifications

    Calibration Manual Specifications ±20.99 mA ±21.0 mA - ±99.9 mA Current Ranges 0.01 mA 0.1 mA Resolution Accuracy 0.2 % reading ±5 digits 20.99 mA range 1 % reading ±5 digits 99.9 mA range ±99.9 mA Maximum Reading < 0.20 mA Influence of Earth’s Field 2 AA 1.5 V Alkaline, IEC LR6 Battery...
  • Page 13: Getting Acquainted With The Meter

    Milliamp Process Clamp Meter Getting Acquainted with the Meter Getting Acquainted with the Meter Figure 1 shows the Meter functions and features. ege01a.eps Number Description Turns the Meter on and off. When the Meter is in sleep mode, press this button to wake it up. Captures and holds the current reading Removes interference and zeros the display Measurement spotlight LED button...
  • Page 14: Maintenance

    Calibration Manual Maintenance XWWarning To avoid possible electric shock or personal injury, repairs or servicing not covered in this manual should be performed only by qualified personnel. Cleaning the Meter XWWarning To avoid electrical shock, remove any input signals before cleaning.
  • Page 15 Milliamp Process Clamp Meter Maintenance ege02.eps Figure 2. Changing the Batteries...
  • Page 16: Performance Tests

    The following tests are used to verify the functions of the Meter. If the Meter fails any of the verification tests, repair is necessary. For service, see Contacting Fluke. Required Equipment Required equipment for the performance tests is listed in Table 2. If the recommended models are not available, equipment with equivalent specifications may be used.
  • Page 17: Testing The Display

    Milliamp Process Clamp Meter Performance Tests Testing the Display 1. Press and hold Awhile powering on the Meter. 2. Compare the Meter display to Figure 3. 3. Check all display segments for clarity and contrast. fdp1.eps Figure 3. Display Test Display Hold Test 1.
  • Page 18 7. Complete steps 2-4 for each calibrator output setting in Table 3. 8. If the Meter fails to meet any of the Display Reading Limits, it requires calibration adjustment, or repair. See Calibration Adjustment or Contacting Fluke. Table 3. Accuracy Tests...
  • Page 19: Calibration Adjustment

    Milliamp Process Clamp Meter Calibration Adjustment Calibration Adjustment Adjustment Subroutines The Meter features closed-case calibration adjustment using a known reference source. The Meter measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. There are four adjustment subroutines in the Meter adjustment procedure: Low Range (±20 mA) High range (±100 mA) Temperature...
  • Page 20: Calibration Error Messages

    Calibration Manual Calibration Button fdp3.eps Figure 4. Accessing the Calibration Button Calibration Error Messages Table 4 lists the calibration error messages that can be shown on the Meter display. The suggested actions to eliminate messages are also listed. Table 4. Error Messages Error Cause of Error Suggested Action...
  • Page 21: Temperature Adjustment Procedure

    Milliamp Process Clamp Meter Calibration Adjustment Temperature Adjustment Procedure 1. Press C for >1 second until t23 appears on the percentage display. 2. Wait at least 60 seconds for the internal temperature to balance. 3. Press H to adjust the temperature. Low Range Adjustment Procedure 1.
  • Page 22: Phase Adjustment Procedure

    Calibration Manual Phase Adjustment Procedure 1. Clamp the Meter’s current clamp around an insulated 18 gauge, copper wire, with 6-inch diameter. Current flow should be in the direction of the arrow on the current clamp. 2. Press C >1 second until the percentage display indicates the Meter’s currently saved phase value.
  • Page 23: User Replaceable Parts

    Milliamp Process Clamp Meter User Replaceable Parts User Replaceable Parts Table 5 and Figure 6 list all user replaceable parts. Table 5. Replaceable Parts Item ID Description Decal 2723063 Case top assembly 2720362 (does not include decal, order separately) Keypad 2723056 Jaw assembly (Includes cable) 2722971...
  • Page 24 Calibration Manual fdp4.eps Figure 6. Replaceable Parts...