Data set names
DATA SET NAME Entry of a data set name
This function is used for entering the data set name.
Operation:
– Highlight the left function group in the center function
table using JC.
– Select the required data set number with M.
– Simultaneously press N and O allocated to the
function DATA SET NAME.
You will see the following display:
The cursor flashes at the entry position of the inverted
field of the menu.
Krautkramer USN 52R/USN 52L
In the two lower lines you will see letters, numbers and
characters which you can use for the data set name.
To enter the required text, start by selecting the entry
position in each case, and then select the character
which should be inserted at this position.
The entry position and character are always marked by
a flashing cursor. You can enter all required characters
in any chosen order.
– Press the keys T or V in order to change the entry
position.
– Press the keys N or O next to the corresponding
character string in order to select a character. The
flashing cursor changes the string automatically. The
highlighted character is immediately inserted at the
entry position.
– Finally, press the key X to store the entries.
The data set name is now stored together with the data
set number and also appears in the function group
MEM in the first operating level.
Issue 05, 02/00
Operation
5-51