Operating the DGS
H Note:
If the DGS function is switched on while you vary a
parameter that affects the stored reference echo, or
when you use another probe, the following message is
displayed:
"Clear REF and confirm with HOME"
This text remains blinking on the display for a few sec-
onds. If you press the F key during this time, the refer-
ence echo is cleared and the DGS function is switched
off.
DGS-26

3.4 Switching off the DGS mode

If you switch off the DGS mode, the USN 52 R option
DGS automatically sets the gain to the total of refer-
ence gain and incremental gain.
– Switch the DGS mode off in the DGS function:
N or O
After switching off the DGS mode, the reference echo
remains stored. You can clear it by pressing M be-
longing to the GAIN function.
Issue 02, 03/99
Switching off the DGS mode
Krautkramer USN 52R/USN 52L Option DGS