Setting the flaw location calculation
O-DIAM Test object diameter
You need the function O-DIAM if you are dealing with
circular curved surfaces, e.g. when testing longitudi-
nally welded tubes. To enable the instrument to make
the corresponding correction of (reduced) projection
distance and depth, you have to enter the outside diam-
eter of the test object at this point.
If you want to make the flaw position calculation for
plane-parallel test objects, the function O-DIAM has to
be set to infinite.
If O-DIAM is set to a value other than infinite, the fol-
lowing status indicator will appear on the screen:
Operation:
– Use M to set the outside diameter of the test
object in O-DIAM, or select infinite if you are testing
plane-parallel test objects.
H Note
You can select infinite directly by simultaneously
pressing N and O .
Krautkramer USN 52R/USN 52L

Displaying the measured values

The determined values are displayed in the measure-
ment line:
S =
Sound path
D =
Depth
PR = Reduced projection distance, this is displayed
when X-VALUE is not 0.
Otherwise this will read:
P =
Projection distance.
– With U select the value which is to be indicated in
enlarged figures on the display.
H Note
No values for P, PR and D are displayed if, according to
geometrical conditions, the sound beam does not hit
the inner surface of the workpiece.