6. Enter limit values

After you have entered in the time and date, the first
programmable limit value for systolic blood pressure
blinks in the display.
Press the button
5 mmHg steps. The adjustment range for systolic
blood pressure lies between 70-250 mmHg.
Press the button
move to programming the diastolic limit values.
Press the button
5 mmHg steps. The adjusment range for diastolic
blood pressure lies between 40-245 mmHg.
Press the button
Repeat the last five instructions to enter the
value limits for further users (2 + 3).
Press the button
the programming of limit values.
The programming of limit values into each of the
3 memory slots ("no. 1" to "no. 3") can also be
carried out at a later time.
, to raise the value in
SET
, to save this setting and
MODE
SET
, to raise the value in
, to save the setting.
MODE
START / STOP
9
, to discontinue