Using the Current Output Functions

The meter provides steady, stepped, and ramped current
output for testing 0-20 mA and 4-20 mA current loops.
You can choose source mode, in which the meter
supplies the current, or simulate mode, in which the
meter regulates current in an externally-powered current
loop.

Source Mode

Source mode is selected automatically by inserting the
test leads into the SOURCE
Figure 7. Use source mode whenever you need to supply
20
and
jacks as shown in
current into a passive circuit such as a current loop with
no loop supply. Source mode depletes the battery faster
than simulate mode, so use simulate mode whenever
possible.
The display looks the same in source and simulate
modes. The way to tell which mode is in use is to see
which pair of output jacks is in use.