Accuracy Tests For Volts, Current, , E, And G Test Functions; Temperature Test Input And Display; Performance Test Steps - Fluke 287 Calibration Manual

True-rms digital multimeters
Hide thumbs Also See for 287:
Table of Contents

Advertisement

1. Turn the Meter on and position the rotary switch to N and select the temperature
function. Select the degrees C function key.
2. On the Meter, select Offset. Adjust the offset to 000.0 using keypad edit buttons.
3. Close the Offset window.
4. Set the 5520A for K-type thermocouple and an output of 0 C.
5. Perform the measurement steps in Table 3.
Input
0.0 C
100.0 C
1000.0 C
Accuracy Tests for Volts, Current,
To verify the accuracy of Meter functions, do the following:
1. Connect the Calibrator Normal terminals to the Meter's V and COM input terminals.
2. Turn on the Meter.
3. Turn the rotary switch to N.
The DC mV measurements in Table 4 are subject to offsets caused by
thermals due to temperature gradients between connecting materials. After
connecting the calibrator to the Meter terminals, allow time (could be
several minutes) for the thermals to dissipate. When readings are stable,
perform a REL and record the Meter reading.
4. Apply the Input Level and frequency for step 1 of Table 4.
5. Compare the Meter's display reading with the Display Limits in Table 4.
6. If the display reading falls outside of the Display Reading Limits shown in Table 4,
the Meter does not meet specification and requires adjustment or repair.
7. Complete the remaining test steps for each function listed in Table 4.
Step
Function
1.
DC mV
2.
DC mV
3.
DC mV
4.
DC mV
Table 3. Temperature Test Input and Display
Lower Limit
-1.0 C
98.0 C
989.0 C
,
E
, and
Note
Table 4. 287/289 Performance Test Steps
Range
Input Level
50.000 mV
0 mV
REL Offsets
[2]
50.000 mV
0 mV
50.000 mV
0.025 mV
50.000 mV
-0.025 mV
Display
Upper Limit
G
Test Functions
Display Reading
Frequency
Lower
or Model
Limit
Allow reading to stabilize
-0.020
[2]
0.005
[2]
-0.045
True-rms Digital Multimeters
Performance Tests
1.0 C
102.0 C
1011.0 C
[1]
Upper
Limit
0.020
0.045
-0.005
17