4.
Ensure the battery compartment gasket () is
properly installed around the outside edge of the
battery compartment barrier.
5.
Replace the battery door by aligning the battery
compartment barrier with battery compartment.
6.
Secure the door with the six Phillips-head screws.

How to Replace the Fuses

Referring to Figure 11, examine or replace the Meter's
fuses as follows:
1.
Turn the rotary switch to OFF and remove the test
leads from the terminals
2.
Refer to step 2 under the How to Replace the
Batteries section above to remove the battery door.
3.
Remove the fuse compartment seal () from the
fuse compartment.
4.
Gently lift out the fuse compartment door () from
the fuse compartment.
5.
Remove the fuse by gently prying one end loose,
then sliding the fuse out of its bracket ().
6.
Install ONLY specified replacement fuses with the
amperage, voltage, and speed ratings shown in
Table 9. The 440-mA fuse is shorter than the 10-A
fuse. For correct placement of each fuse, note the
marking on the printed circuit board under each fuse.
7.
Replace the fuse compartment door by aligning the
arrow on the fuse door with the arrow on the case
bottom and lowering the door into the fuse
compartment.
8.
Replace the fuse compartment seal by aligning the
tab on the seal with the outline on the case bottom.
Ensure the seal () is properly seated.
9.
Refer to steps four through six under the Replacing
the Batteries section above to reinstall the battery
door.

Service and Parts

If the Meter fails, check the battery and fuses. Review this
manual to verify proper use of the Meter.
Replacement parts and accessories are shown in Table 9
and Figure 12.
To order parts and accessories, refer to "How to Contact
Fluke".
Digital Multimeters