2
Operation and Features
Measuring DC Current
32
SCPI commands
The following example shows how to make an AC voltage
measurement using SCPI commands.
MEASure[:VOLTage]:AC?
The DC current measurement function has the following
features:
• Three ranges to select: 10 mA, 100 mA, 1 A and 2 A or,
auto- range.
• Input protection fuse is 2 A, voltage rating 250 V on all
ranges.
Make the connection as shown below.
Agilent Measurement Manager operation
Select the DCI function and desired range. A suitable range
should be selected to give the best measurement resolution.
The reading is displayed and updated continuously.
–
+
DC Current Source