Operation and Features
2
The INITiate command changes the state of the triggering
to wait- for- trigger state. Measurements will begin when the
specified trigger conditions are satisfied.
Star Triggering
The star trigger can only be applied when the U2741A is
connected into the U2781A modular instrument chassis. It is
used to trigger multiple modular units in the chassis.
SCPI command
TRIGger:SOURce STRG
Synchronization Status
This is to configure the synchronization of multiple units of
U2741A (slave only) when used in the U2781A modular
instrument chassis. Only one master can be assigned at a
time.
SCPI command
CONFigure:SSI {NONE|SLAVe}
Refer to the Agilent U2741A Programmer's Reference for
complete description and syntax for these commands.
U2741A User's Guide