13.2

LR530 flanged encapsulated antenna

LR530
- - - - Sensor reference point
Dimensions in mm(inch)
Process connection type A mm
flange DN25 PN6, raised
face, Form B1,
EN1092-1, DIN2501 /
316/316L
flange DN50 PN6, raised
face, Form B1,
EN1092-1, DIN2501 /
316/316L
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA
A with
Bea
(inch)
m
thermal
an‐
exten‐
gle
sion
201.7
221.7
10°
(7.94)
(8.73)
216.1
268.1
(8.51)
(10.56)
13.2 LR530 flanged encapsulated antenna
LR530 with thermal extension
 
Recommen‐
Flange
ded max.
O.D.
range [m
[mm
(ft)]
(inch)]
20 (65.6)
100
(3.94)
30 (98.4)
140
(5.51)
Dimension drawings
Bolt hole
Bolt
No.
circle Ø
hole
of
[mm
Ø [mm
bolt
(inch)]
(inch)]
holes
75 (2.95)
11
4
(0.43)
110 (4.33)
14
4
(0.55)
Flange
thick‐
ness
[mm
(inch)]
16
(0.63)
20
(0.79)
189