Siemens SITRANS FMS500 Operating Instructions Manual page 47

Electromagnetic flowmeters
Table of Contents

Advertisement

Version
Corrosive category
Pressure drop
Test pressure
Mechanical load (vibration) - compact version
 
With compact transmitter FMT020
Mechanical load (vibration) - remote version
 
Sensor
 
Transmitter FMT020
Process fluid temperature
 
NBR
 
EPDM
EMC
1
Maximum operating pressure decreases with increasing operating temperature
For PTB K7.2 DN 15 to 1200
2
Table 8-4
Design
Version
Housing and flange material
Measuring pipe
Electrodes
Grounding electrodes (standard)
Terminal box
SITRANS FMS500
Operating Instructions, 01/2024, A5E52677520-AA
FMS500
C4 (C5 optional) according to ISO 12944-2
DN 15 and 25 (½" and 1"):
Max. 20 mbar (0.29 psi) at 1 m/s (3 ft/s)
DN 40 to 300 (1½" to 12"):
Max. 25 mbar (0.36 psi) at 3 m/s (10ft/s)
DN 350 to 1200 (14" to 48"):
Insignificant
1.5 x PN (where applicable)
 
Vibration, sinusoidal according to IEC 60068-2-6: 1 g peak
Vibration broad-band random, according to IEC
60068-2-64: 1.54 g rms
 
Vibration, sinusoidal according to IEC 60068-2-6: 2 g peak
Vibration broad-band random, according to IEC
60068-2-64: 1.54 g rms
Vibration, sinusoidal according to IEC 60068-2-6: 0.7 g
peak
Vibration broad-band random, according to IEC
60068-2-64: 1.54 g rms
 
-10 to +70 °C (14 to 158 °F)
-10 to +70 °C (14 to 158 °F)
EMC 2014/30/EU
FMS500
Carbon steel, with corrosion resistant coating of corrosive cat‐
egory C4 or C5, according to ISO 12944-2
AISI 304 (1.4301)
Hastelloy
Hastelloy
Polycarbonate
Technical specifications
8.1 FMS500
47