Connecting
5.2 Connecting the sensor
4. Cut off the black wire on the coil cable. Fit coil and electrode cable through the cable glands.
5. Connect coil cable
blocks. Tighten all terminal screws with maximum torque: 0.5 Nm.
6. Connect the coil and electrode cable shields and grounding cables to the terminal board by
using grounding lugs.
30
①
②
and electrode cable
in the corresponding terminals on the terminal
Operating Instructions, 01/2024, A5E52677520-AA