Troubleshooting
Picture displayed
Screen
LED
Picture displayed
Green
Yellow
Yellow (flashing)
Red
See also
Switching on the monitor and PC (Page 31)
40
Cause
No error, correct operating status
Lamp warm-up period:
Setting is active, and the display is in the warm-
up period.
Display has reached an initial critical temperature
level.
Lamp warm-up period:
Setting is active, and the warm-up period has
expired without the display having reached the
stabilized luminance level.
Internal error
Remedy
-
Wait for the warm-up period
to expire.
The LED turns green when
the lamp has reached the
stabilized luminance level.
Select a lower luminance
level for standard operation.
Check the ventilation and
improve these conditions if
necessary.
Inform servicing department
Inform servicing department