High Speed Mode - Fluke 8505A Instruction Manual

Hide thumbs Also See for 8505A:
Table of Contents

Advertisement

8505A/8506A
Table
2A~2,
Programming
Instructions (85QSA,
8S08A)
The
following discussion
relates
remote
operation features
available with
the
8505A and
the
8506A
to existing
documentation
for
remote
operation
of
the
8500
series
multimeters.
The
additional features for
the
8505A and
the
8506A
are
presented
in
this
discussion
in
the
same
sequence
as
they
would appear
in
the
"Programming
Instructions" table (Table 2A-1).
Most
of
the items
documented
in
this
table
supplement
features available with the
8502A
and 8502A/AT.
Incompatibilities
have been
kept
to
a
minimum,
Therefore,
programs designed
for
the
8502A
are generally
compatible
with the
8505A and
the
8506A,
and
8502A/AT programs
are generally
compatible
with the
8506A.
The
few
areas
that are
not
compatible
are
briefly
described below. Refer
to
appropriate areas
in
this
table
for
a
more
detailed
description.
1
.
High
Speed
Mode
(!):
Some
High
Speed
mode
scaling
factors
have been changed.
Also,
use
of the
reset
command
(*)
causes
the multimeter
to
both
exit
High Speed
mode
and perform
a
normal
reset.
2.
Range Commands:
Full
scale points
and
autoranging
points
have
been changed
in
several instances.
3.
Store
Zero
Commands
(K0,
K1
):
In
the
8505A/8506A,
the
temporary zero
correction
values are
set
to
0
when
the instrument
is
reset
or
powered
up
{similar to
the
8502A
without the
calibration
memory
option).
In
the
8502A
with the calibration
memory
option, resetting
the instrument
does
NOT
clear the stored zeros.
Interruption
of
input
power
could
affect
Calibration
Memory
entries
when
the
multimeter
is
in
Calibration
mode.
Do
not cycle input
power
to
the multimeter
when
Calibration
mode
is
activated.
If
power
is
ON,
verify that
the
AVG/(CAL)
annunciator
is
not
flashing
before cycling
power
to
OFF.
From
the
remote,
Calibration
mode
status
can
be
verified
with the
G5 command
(response
of
0
=
mode
off,
response
of
1
=
mode
on),
if
power
is
OFF,
verify that
the
rear
panel
Calibration
switch
Is
off
before cycling
power
to
ON.
Both
the
8505A
and
the
8506A assume
the
same
configuration
at
power up
as
that
described
for
8502A,
with the
following exceptions:
1.
The
Average
mode
is
disabled (OO).
2.
Calibration
Memory
factors are
retained
at
alt
times.
These
factors
include
zero corrections
for
each
range
in
dc
volts
and
ohms
functions,
gain corrections
for
each range
in
each
function,
and
the
calibration
date
(or
instrument
identification)
number.
3.
External
Reference
at
power up
or reset
is
used
as
temporary
storage
for
the multimeter software
version
number.
The
GNR
command
can
then be
used
to
recall this
number.
Any
use
of
the
X
command
subsequent
to
power-up
or
reset
replaces
this
number
with the
value applied
at
the external reference
inputs.
4.
Zero
mode
is
enabled
with
all
temporary
zero correction values
set
to
0.
HIGH SPEED
MODE
Selection
of the
High
Speed
mode
sets
the following conditions:
1.
The
"!"
command
both enters the
High
Speed
mode
and
triggers
a reading.
2.
Use
of
any
character
(or
bit
pattern)
other than "?"
causes
the
multimeter
to exit
the
High
Speed mode. Use
of
causes
the multimeter
to
both
exit
the
High
Speed
mode
and
perform a normal
reset.
The
character
used
to exit
High
Speed
mode
must be
sent by
itself.
Any commands
to
be executed
after
exiting
High
Speed
mode
must be
sent
in
a separate transmission
(i.e.,
in
a
separate statement
in
the instrument
controller
program).
3.
Selection of
High
Speed
mode
specifies the
binary output
format.
Any
previously selected
output format
is
restored
once
High
Speed
mode
is
exited.
4.
The
multimeter
front
panel
display
is
blank while
High
Speed
mode
is
on.
The
previously selected display
mode
is
restored
once
High
Speed
mode
is
exited.
2A-22

Advertisement

Table of Contents
loading

Table of Contents