Wiring
3
Wiring
3.1

Tantalum wiring

DANGER!
Make sure the hazardous area specified on the sensor approval tag is suitable for the
environment in which the sensor will be installed. Failure to comply with the requirements for
intrinsic safety in a hazardous area could result in an explosion.
CAUTION!
Fully close and tighten all housing covers and conduit openings. Improperly sealed housings
can expose electronics to moisture, which can cause measurement error or flowmeter failure.
Inspect and grease all gaskets and O-rings.
3.1.1
Connect the 9-wire cable
1.
2.
6
A 9-wire cable is limited to 1000 feet (305 meters) between the sensor and the
transmitter.
If you are using the Smart Meter Verification (SMV) application, the 9-wire cable is
limited to 60 feet (18 meters) maximum.
Use Teflon
jacketed cable for SMV applications.
Prepare and install the cable according to the instructions in the Micro Motion 9-
Wire Flowmeter Cable Preparation and Installation Guide.
Insert the stripped ends of the individual wires into the terminal blocks of the
junction box. Ensure that no bare wires remain exposed.
Micro Motion TA Series