GE 325 Product Manual page 42

Maternal/fetal simulator
Table of Contents

Advertisement

SECTION 4
4.
Set the UA Level switch to O. (The units represent mmHg
or relative units for the modes IUP and TOCO,
respectively.)
5.
Briefly press the monitor's UA Reference pushbutton to
set the UA baseline at the monitor's default setting.
NOTE: For the IUP mode, the baseline is always set at 0
mmHg. For the TOCO mode, most monitors default to 10
relative units. However, some monitors allow the default
baseline to be configured to another value. Refer to the
respective operator's manual for more information.
6.
Set the UA Level switch to the desired pressure level.
(Refer to Section 3, Controls, Indicators, and Connectors,
for information about selections.)
M
LEVEL mmHg
CMR
Power
μπω
LEVEL mmHg
CMR
— MANUAL
~ RAP
100
50
10
n
o
4-12 + Operation