Chapter 3 Inspection
5.
Press again the power switch to turn the light source OFF.
6.
Connect an endoscope to the light source.
Refer to Section 4.2, "Connection of an endoscope" on
page 50 for connecting an endoscope.
/.
Press the power switch of the light source. The power indicator lights up
(see Figure 3.3).
Power
indicator
Power switch
Figure 3.3
8.
Confirm that air is exhausted by holding your hand in front of the ventilation
grills on the rear panel.
WARNING
If the exhaust does not emit air, do not use the light source
and contact Olympus.
O
If the power fails to come ON
If the power fails to come ON, turn the light source OFF. Then, confirm that the
power cord is connected firmly and that the lamp cover is firmly closed. Then,
turn the light source ON again. If the power fails to come ON, replace the fuses
with new ones as described in Section 5.2, "Fuse replacement" on page 72. If
the power still fails to come ON, contact Olympus.
28
OLYMPUS
EVIS EXERA II XENON LIGHT SOURCE CLV-180
Chapter 3 Inspection
O
If the emergency lamp indicator lights up
If the emergency lamp indicator lights up, turn the light source OFF and then ON
again and try to ignite the examination lamp again. If the emergency lamp
indicator still lights up, replace the examination lamp with a new one as
described in Section 5.1, "Replacement of the examination (xenon) lamp" on
page 64. Ifthe emergency lamp indicator still lights up, contact Olympus.
Pressing the lamp button when the examination lamp is OFF
or turning the light source ON when the auto ignition is active
automatically tries to ignite the examination lamp for
5 seconds. If ignition fails, the light source automatically
switches to the emergency lamp and the emergency lamp
indicator lights up.
O
If the emergency lamp indicator blinks
If the emergency lamp indicator blinks, the installed emergency lamp is not
working properly. Contact Olympus.
O
If the indicators on the control panel blink
If the indicators on the control panel blink, the light source is not working
properly. Contact Olympus.