Chapter 4 Inspection

4.1

Inspection of the power supply

ELECTROSURGICAL UNIT PSD-30
Before each case, inspect this electrosurgical unit as
instructed below. Inspect other equipment to be used with
this electrosurgical unit as instructed in their respective
instruction manuals. Should the slightest irregularity be
suspected, do not use the electrosurgical unit and see
Chapter 7, "Troubleshooting". If the irregularity is still
suspected, after consulting Chapter 7, contact Olympus.
Damage or irregularity may compromise patient or user
safety and may result in more severe equipment damage.
Prepare this electrosurgical unit and other equipment (shown in the "System
chart" in the Appendix) to be used with this electrosurgical unit for each particular
case. Refer to the respective instruction manual for each piece of equipment.
If a sound is not heard after turning on the power switch, do
not perform high-frequency cauterization.
The high-frequency output may not be noticed and patient
injury, bleeding and/or perforation could result.
Press the power switch to turn ON the electrosurgical unit. Confirm that a sound
is heard. At the same time all the indicators turn on and then turn off, and then
the coagulation mode indicators, warning indicators and cut mode indicators
should light alternately (see Figure 4.2). Then the power switch illuminates (see
Figure 4.1) and the cut mode indicators and coagulation mode indicators should
flash sequentially.
Chapter 4 Inspection
35