Inspection of the high intensity mode
NOTE
The high intensity mode is not available when no endoscope
is connected to the light source. The function is available only
when a compatible endoscope is connected.
1.
Press the AUTOJMAN.
brightness button and confirm that the "MAN."
indicator lights (see Figure 3.11).
Press the brightness control buttons to set the brightness level to O.
Connect an endoscope compatible with high intensity mode to the output
socket on the light source.
Press the intensity mode button to change the normal mode to high intensity
mode (see Figure 3.11).
Confirm that the high intensity mode indicator lights and the light emitted
from the endoscope's distal end increases compared to the normal mode.
6.
Press the intensity mode button to switch to the normal intensity mode.
/.
Confirm that the high intensity mode indicator goes out and the light emitted
10.
11.
12.
from the endoscope's distal end decreases compared to the high intensity
mode.
Press the intensity mode button again and confirm that the high intensity
mode is restored.
Disconnect the endoscope. Confirm that the high intensity modg indicator
still lights.
Turn the light source OFF.and turn it ON again. Confirm that the high
intensity mode indicator lights.
Connect the endoscope compatible with the high intensity mode to the
output socket on the light source again. Confirm that the high intensity mode
indicator still lights.
Press the AUTO/MAN.
brightness button and confirm that the "AUTO"
indicator lights.