Chapter 6 Compatible Reprocessing Methods

Chapter 6 Compatible Reprocessing
6.1

Compatibility summary

88
Methods
The endoscope and accessories are compatible with several methods of
reprocessing. For information concerning the applicable reprocessing methods
and parameters, refer to Section 6.2, "List of compatible methods". Additionally,
some reprocessing methods may cause degradation of medical devices that
shortens product life. For information concerning degradation of medical devices
from reprocessing and its number of times, refer to Section 6.13, "Signs of
degradation from reprocessing and its number of times".
Follow the policies at your local institution when choosing which methods to
employ.
The methods listed as "compatible" in Table 6.1 through 6.5
are compatible for routine use only when used according to
the manufacturer's instructions. Repeated use and
reprocessing of endoscopes and accessories lead to gradual
wear and tear. Furthermore, reprocessing methods that
employ higher temperatures and more caustic/corrosive
materials may lead to faster deterioration. In general,
sterilization processes are harsher on equipment than
disinfection processes. Before each patient procedure,
inspect the endoscope and accessories for damage,
according to the instructions described in this manual and its
companion "OPERATION MANUAL".
The instructions provided in this manual regarding
compatible reprocessing methods are not valid for Olympus
devices repaired by a non-Olympus facility. Olympus repairs
devices to manufacturer's specifications using original
equipment manufacturer's (OEM) materials. The use of
non-OEM materials to repair an Olympus device may affect
the material compatibility and reprocessing efficacy of the
device with certain reprocessing chemicals or methods. In
the event that your device has been repaired by a
non-Olympus facility, contact the repair facility for instructions
regarding compatible reprocessing methods.
The ultrasonic cable is not waterproof. Never immerse it in
disinfectant solution or any other fluids.
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180