Front Panel Features (Figure 342) - HP 4262A Operating And Service Manual

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Section III
Paragraphs 3-49 to 3-51
Model 4262A
349.
OPTION 004- COMPARATOR.
3-50.
The 4262A Option 004 (shown in Figure
3-43) provides:
(a)
HIGH and LOW limits setting for com¬
parison of LCR and D/Q measured data.
(b)
LED visual decision output lamps display
of results of HIGH and LOW limit com¬
parisons.
(c)
TTL outputs and relay outputs for HIGH,
IN, and LOW decision outputs.
3-51.
Front Panel Features (Figure 342).
(1)
LCR LIMIT Switch: Two four-digit
switches provide HIGH and LOW limit
values with which measured LCR value is
compared. Setting range is from 0000 to
1999.
(2)
LCR Decision Output Lamp: Results of
comparison are indicated by LED lamps
as follows;
HIGH: (measured value £ High limit)
IN: (Low limit £ measured value
< High limit)
LOW: (measured value < Low limit)
Figure 3-42. Front Panel Features
(4)
D/Q LIMIT CHECK Switch: While this
switch is depressed, HIGH and LOW limit
values set by D/Q LIMIT switches (6) are
displayed in LCR and D/Q displays. During
this period, three D/Q lamps of decision
outputs are lit.
(5)
D/Q Decision Output Lamp: Results of
comparison is indicated by LED lamps
as follows:
(3) LCR LIMIT CHECK Switch: While this
switch is depressed, HIGH and LOW limit
values set by LCR LIMIT switches (1) are
displayed in LCR and D/Q displays. During
this period,
three LCR decision output
lamps are lit.
Comparator must be en¬
abled display limits.
HIGH .-(measured value ^ High limit)
IN: (Low limit £ measured value
< High limit)
LOW: (measured value < Low limit)
(6) D/Q
LIMIT
Switch:
Two
four-digit
switches provide HIGH and LOW limit
values with which measured D/Q value is
compared. Setting range is from 0000 to
1999.