Using Current Bias (for inductors),
1.
Connect dc power supply as shown below:
Note
DC power supply should be floated from ground.
If cable between low terminals of 4262A and
dc ammeter
2. Set 4262A controls as follows:
DC BIAS........._......-...... EXT
FUNCTION...
L
CIRCUIT MODE. ............... .PRL or SER
LCR RANGE ... . .MANUAL
Other controls. ..
any settings
Note
First, determine appropriate range by connect¬
ing sample with no dc bias current applied.
Then hold the range.
3.
Recommended inductance ranges and maximum bias currents are:
Range (at 120Hz)
1000
uH
10.00 mH
100.0 mH
1000 mH
10.00 H
100.0 H
CIRCUIT MODE
SER
PARA
Maximum Bias
Current
40mA
36mA
13mA
40mA
36mA
13mA
In IkHz(lOkHz) measurement, multiply range value by 1/10 (1/100).
CAUTION
DC BIAS OVER +40 VOLTS MUST NOT BE AP¬
PLIED TO EXTERNAL DC BIAS INPUT CON¬
NECTOR,
Figure 3-9, External DC Bias Circuit (Sheet 3 of 3).