HP 4262A Operating And Service Manual page 11

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262A
Table 1-1. Specifications (Sheet 4 of 4).
Section I
Table 1-1
OPTIONS
Option 001: Simultaneous BCD output of LCR
and DQ data (positive true). Max. sink
current 16mA. Mating connector (P/N
1251-0086). (Alternate BCD output
of LCR and DQ data selectable by
switch on internal board).
Option 004: Digital comparator (can
not be
used with OPT 101). Compares meas¬
ured value with high and low limit set¬
tings for LCR or DQ and provides
HIGH, IN, LOW comparison outputs.
Limit setting range: 0000
- 1999
for each
limit switch.
Comparison output: Visual, relay contact,
and TTL level.
Visual: 3 LED's indicate HIGH(red), IN
(green), or LOW (red).
Relay contacts:
SPST
contacts to circuit common
for each HIGH, IN and LOW output.
TTL level:
Open collector circuits to high level
(open) for each HIGH, IN and LOW
outputs (fanout max. 30mA).
Option 101: HP-IB data output & remote con¬
trol.
Remotely controllable functions:
Function (L, C, R/ESR, a LCR)
Loss
(D, Q)
LCR range
DQ range
Circuit mode
Test frequency
&
level
Trigger
Self test
Data output: C - D/Q, L - D/Q, R/ESR
Internal function allowable subsets:
SHI, AH1, T5, L4, RL1, DC1, SRI
and DTI.
Data output format: Either
of two formats
may be selected. Switchable at rear
panel (no + sign outputs).
Format A.
SFFT±N.NNNE+NN, SF±N.NN(6r)@
Format B.
SFFT±N.NNNE±NN(CR)@
SF±N.NN@)@)
SFFT
L-Test Frequency
—Measurement Equivalent Circuit
^-Measurement Status
SF
L-Loss measurement D or Q
Measurement status
Option 010: 100Hz test frequency instead of
120Hz.
ACCESSORIES AVAILABLE
16061A: Test fixture, direct coupled, 5-terminal
Two kinds of inserts are included for
components with either axial or radial
leads. Usable on all ranges of 4262A.
16062A: Test cable with alligator clips, 4-termi-
nal. Useable for low impedance meas¬
urements. Measurement range at 1kHz
is L
2H, C > 200nF and R <, 10ka
[For L and C measurements, these
ranges increase by xlO at 120 (100)Hz
and
decrease
by
same
factor
at
10kHz].
16063A: Test cable with alligator clips, 3-termi-
nal. Useable for high impedance meas¬
urements. Measurement range at 1kHz
is L ^ 3mH, C £ 10m F and R 2:200«.
[For L and C measurement, these
ranges increase by xlO at 120(100)Hz
and
decrease
by
same
factor
at
10kHz].
1
-

Advertisement

Table of Contents
loading