Counter n A, Counter n B, and Counter n Z Signals

Counter n Up_Down Signal

Counter n HW Arm Signal

© National Instruments Corporation
ai/StartTrigger
PXI_STAR
In addition, Counter 1 Internal Output, Counter 1 Gate, Counter 1 Source,
or Counter 0 Gate can be routed to Counter 0 Aux. Counter 0 Internal
Output, Counter 0 Gate, Counter 0 Source, or Counter 1 Gate can be routed
to Counter 1 Aux.
Some of these options may not be available in some driver software.
Counter n B can control the direction of counting in edge counting
applications.
Use the A, B, and Z inputs to each counter when measuring quadrature
encoders or measuring two pulse encoders.

Routing Signals to A, B, and Z Counter Inputs

Each counter has independent input selectors for each of the A, B, and Z
inputs. Any of the following signals can be routed to each input.
RTSI <0..7>
Input PFI <0..5>
PXI_STAR

Routing Counter n Z Signal to an Output Terminal

You can route Counter n Z out to RTSI <0..7>.
Counter n Up_Down is another name for the Counter n B signal.
The Counter n HW Arm signal enables a counter to begin an input or output
function.
To begin any counter input or output function, you must first enable, or arm,
the counter. In some applications, such as buffered semi-period
measurement, the counter begins counting when it is armed. In other
applications, such as single pulse-width measurement, the counter begins
7-29
Chapter 7
Counters
NI 6238/6239 User Manual