914X Field Metrology Wells
Specifications
Base Unit Specifications
Weight
Power Requirements
System Fuse Ratings
4–20 mA Fuse (-P
model only)
Computer Interface
Safety
Table 3
-P Option Specifications
-P Specifications
Built-in Reference Thermometer Readout
Accuracy
(4-Wire Reference Probe)
Reference Resistance Range
Reference Resistance Accuracy‡
Reference Characterizations
Reference Measurement Capability
Reference Probe Connection
Built-in RTD Thermometer Readout Accuracy
RTD Resistance Range
Resistance Accuracy
RTD Characterizations
RTD Measurement Capability
RTD Connection
14
Shop for Fluke products online at:
9142
8.16 kg (18 lbs)
100 V to 115 V (± 10 %)
50/60 Hz, 635 W
230 V (± 10 %) 50/60 Hz,
575 W
115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
RS-232 and 9930 Interface-it control software included
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04
MyFlukeStore
www.
9143
7.3 kg (16 lbs)
100 V to 115 V (± 10 %), 50/60 Hz, 1400 W
230 V (± 10%), 50/60 Hz, 1800 W
115 V: 15 A F 250 V
230 V: 10 A F 250 V
50 mA F 250V
± 0.013 °C at -25 °C
± 0.015 °C at 0 °C
± 0.020 °C at 50 °C
± 0.025 °C at 150 °C
± 0.030 °C at 200 °C
± 0.040 °C at 350 °C
± 0.050 °C at 420 °C
± 0.070 °C at 660 °C
0 ohms to 400 ohms
0 ohms to 42 ohms: ±0.0025 ohms
42 ohms to 400 ohms: ±60 ppm of reading
ITS-90, CVD, IEC-751, Resistance
4-wire
6 Pin Din with Infocon Technology
NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C
PT-100 (3926): ± 0.02 °C at 0 °C
PT-100 (JIS): ± 0.02 °C at 0 °C
0 ohms to 400 ohms
0 ohms to 25 ohms: ±0.002 ohms
25 ohms to 400 ohms: ±80 ppm of reading
PT-100 (385),(JIS),(3926), NI-120, Resistance
2-,3-,4-wire RTD w\ Jumpers only
4 terminal input
.com
9144
7.7 kg (17 lbs)