10 Rosemount 2051 Product Certifications - Emerson Rosemount 2051 Series Quick Start Manual

Pressure transmitter and flow meter
Hide thumbs Also See for Rosemount 2051 Series:
Table of Contents

Advertisement

Quick Start Guide
10  Rosemount 2051 product certifications
Rev 1.28
10.1  European Directive information
A copy of the EU Declaration of Conformity can be found at the
end of the Quick Start Guide. The most recent revision of the EU
Declaration of Conformity can be found at Emerson.com.
10.2  Ordinary location certification
As standard, the transmitter has been examined and tested to
determine that the design meets the basic electrical, mechanical,
and fire protection requirements by a Nationally Recognized Test
Laboratory (NRTL) as accredited by the Federal Occupational Safety
and Health Administration (OSHA).
10.3  Hazardous location certification
NOTICE
Device ambient temperature ratings and electrical parameters may
be limited to the levels dictated by the hazardous location certificate
parameters.
10.4  North America
E5 USA Explosionproof (XP) and Dust Ignition-proof (DIP)
Certificate
Standards
Markings
30
2041384
FM 3600: 2022, FM 3615: 2022, FM 3616: 2022,
ANSI/UL 61010-1-2019 Third Edition, ANSI/UL
12.27.01: 2022 (Fourth Edition), ANSI/UL 50E (First
Edition)
XP CL I, DIV I, GP B, C, & D T5;
SEAL NOT REQUIRED
DIP CL II, DIV I, GP E, F, & G; CL III T5;
T5: (-50 °C ≤ Ta ≤ 85 °C)
TYPE 4X, IP 68
OPTIONAL: SINGLE SEAL
September 2023
Rosemount 2051 and 2051CF

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 2051cf series