EMC Information
CAUTION
The FlexScan MH240W requires special precautions regarding EMC and need to be installed, put into service
and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.
Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the FlexScan MH240W. Doing so
may affect the FlexScan MH240W.
The FlexScan MH240W should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary, the equipment or system should be observed to verify normal operation in the configuration
in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts, configuring a
medical system, is responsible that the system complies with the requirements of EN60601-1-2:2001.
Guidance and manufacturer's declaration - electromagnetic emissions
The FlexScan MH240W is intended for use in the electromagnetic environment specified below.
The customer or the user of the FlexScan MH240W should assure that it is used in such an environment.
Emission test
Compliance Electromagnetic environment - guidance
RF emissions
Group 1
EN55011
RF emissions
Class B
EN55011
Harmonic emissions
Class D
EN61000-3-2
Voltage fluctuations /
Complies
flicker emissions
EN61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The FlexScan MH240W is intended for use in the electromagnetic environment specified below.
The customer or the user of the FlexScan MH240W should assure that it is used in such an environment.
Immunity test
EN60601
test level
Electrostatic
±6kV contact
discharge (ESD)
±8kV air
EN61000-4-2
Electrical fast tran-
±2kV
sient / burst
for power supply lines
EN61000-4-4
±1kV
for input/output lines
Surge
±1kV differential mode
EN61000-4-5
±2kV common mode
Voltage dips, short
<5% U
interruptions and
(>95% dip in U
voltage variations on
40% U
power supply input
(60% dip in U
lines
70% U
EN61000-4-11
(30% dip in U
<5% U
(>95% dip for U
Note: U
of the test level.
Power frequency
3A/m
(50/60Hz)
magnetic field
EN61000-4-8
The FlexScan MH240W uses RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause any interference
in nearby electronic equipment.
The FlexScan MH240W is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance level
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV differential mode
±2kV common mode
, for 0.5 cycle
<5% U
T
)
(>95% dip in U
T
, for 5 cycles
40% U
T
)
(60% dip in U
T
, for 25 cycles
70% U
T
)
(30% dip in U
T
, for 5s
<5% U
T
)
(>95% dip for U
T
is the a.c. mains voltage prior to application
T
3A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceram-
ic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
, for 0.5 cycle
Mains power quality should be that of a
T
)
typical commercial or hospital environ-
T
ment. If the user of the FlexScan MH240W
, for 5 cycles
T
requires continued operation during power
)
T
mains interruptions, it is recommended
, for 25 cycles
T
that FlexScan MH240W be powered from
)
T
an uninterruptible power supply or battery.
, for 5s
T
)
T
The power frequency magnetic field should
be measured in the intended installation
location to assure that it is sufficiently low.
51