Advertisement

5. Set the HP 5334B as follows:
CHAN
A
Negative SLOPE
................................................................................
ON (falling edge)
VERIFY: The Counter displays 30 ns
+
10 ns (Fall Time).
6. Record the fall measurement on the Performance Test Record Card, lime 39.
7 . Set the function generator as follows:
Frequency
...............................................................................................................
100 Hz
Amplitude
..............................................................................................................
1 v P-P
Function
.................................................................................................................
Negative Ramp
VERIFY: The Counter displays 8 ms
2
0.6 ms (Fall Time).
8.
Record the fall measurement on the Performance Test Record Card, line 40.
9. Set the function generator as follows:
Function
.................................................................................................................
Positive Ramp
10. Set the HP 5334B as follows:
CHAN
A
Negative SLOPE
.................................................................................
OFF (rising edge)
VERIFY: The Counter displays 8 ms
+
0.6 ms (Rise Time).
1 1 .
Record the rise measurement on the Performance Test Record Card, line 41.
Failure: If the instrument under test does not meet the test specification, consider performing the adjustments in
Section V of this manual as a first step in correcting the problem.