HP 5334B Service Manual page 35

Universal counter
Hide thumbs Also See for 5334B:
Table of Contents

Advertisement

Procedure:
1 . Set the signal generator as follows:
...............................................................................................................
Frequency
100 MHz
..............................................................................................................
Amplitude
50
mV
rms
NOTE
OPTION
060
If the HP
53348
has Option
060
(rear panel inputs), terminate the unused
Channel A input (front or rear) with a
500
load.
2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
PERIOD
A
GATE TIME
.........................................................................................................
1 Second
SENS
.......................................................................................................................
ON
CHAN
A
TRIGISENS control
...........................................................................
fully CW
3. Connect the signal generator output to the HP 5334B Input
A
using a 50 Ohm feedthrough connector as
shown in Figure 4-8.
NOTE
Do not use a 50R feedthrough connector at the Input for Counters with Op-
tion
060.
VERIFY: The Counter displays 10 ns
+.
.000 000 1 ns.
4.
Record the Period
A
measurement on the Performance Test Record Card, line 33.
Failure: If the instrument under test does not meet the test specification, consider performing the adjustments in
Section V of this manual as a first step in correcting the problem.
HP 5334B
-
Service Manual
4-25