Installation
June 2023
Procedure
1. Install four threaded wall anchors according to the dimensions of the gas
2. Place the mounting bracket on to the wall anchors and tighten the mounting nuts.
Related information
Secure the gas chromatograph (GC) to the mounting bracket
3.4 

Mounting the sample conditioning system

There are several sample conditioning systems (SCS) available for the Rosemount 470XA
Gas Chromatograph (GC).
For multiple stream applications, several plate-mounted options are available that can be
mounted to a pole or wall.
It is also possible to use a third-party SCS. A third-party SCS must contain the following
functional components:
2-micron or better particulate filter
Liquid filter/shut-off
Flow control to limit the sample flow to between 20 and 50 cc/min
30
chromatograph (GC). Use the bracket as a guide to locate the anchors correctly
before drilling the holes.
The threads of the anchors should protrude from the wall by 1½ in. (40 mm).
Ensure that the bracket is attached firmly to the wall.
Manual
MS-00809-0100-0470