5. Select Connect on the Chromatograph screen.
6. Click the Ethernet button in the Direct-DHCP row.
7. Enter your login information and press OK.
4.2 
Interacting with the local operator interface
(LOI)
After the firmware has booted up, the LOI will display the Home screen.
To edit data, you must be logged in at the appropriate security level; if you attempt to edit
data without being logged in, the Login screen will appear.
You will be logged off automatically after a period of 15 minutes of inactivity. The LOI will
turn off the back light and return to the Home screen.
Rosemount 470XA
NOTICE
Refer to our
Downloads page
MON2020.
The Connect to GC screen opens.
If the Direct-DHCP row is not visible, go to File → GC Directory and click Add to
create a new entry. Make sure the IP address of the new entry is 192.168.135.100
and name it Direct-DHCP.
The Login screen opens.
The computer connects to the GC. The status bar at the bottom of the Rosemount
MON2020 screen shows the GC name, the alarm status, the mode, and the time and
date of the connected GC.
to download the latest version of Rosemount
Using the Rosemount 470XA
June 2023
67